Entrance surface air kerma to patients during digital radiographic examinations in Tanzania

By: Contributor(s): Material type: TextTextLanguage: English Series: ; Radiation Protection Dosimetry, Volume 200, Issue 3, March 2024Publication details: Mwanza, Tanzania : Catholic University of Health and Allied Sciences [CUHAS-Bugando] : 2023Description: Pages 229–239; Includes ReferencesISSN:
  • Online ISSN 1742-3406
Subject(s): Online resources: Summary: Abstract : The aim of this study was to determine the entrance surface air kerma (ESAK) in adult patients during digital radiography and to evaluate the optimisation potential in five common X-ray examinations in Tanzania. Based on a sample of 240–610 patients, ESAK was estimated using X-ray tube output measurements, patient information and backscatter factors. The results show that the mean ESAK values were higher or comparable to data from the literature. The diagnostic reference values of ESAK for digital radiography were 0.31 mGy (chest PA), 4 mGy (lumbar spine AP), 5.4 mGy (lumbar spine LAT), 3.8 mGy (abdomen AP) and 2.4 mGy (pelvis AP). For computed radiography, the mean ESAK ranges were 0.44–0.57 mGy (thoracic AP), 3.59–3.72 mGy (lumbar spine AP), 6.16–6.35 mGy (lumbar spine LAT), 3.89–3.44 mGy (abdominal AP) and 2.92–3.47 mGy (pelvic AP). In conclusion, high ESAK variations show the potential for optimising protection in digital radiology.
Item type: RESEARCH ARTICLES
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)
Holdings
Item type Current library Status Barcode
RESEARCH ARTICLES MWALIMU NYERERE LEARNING RESOURCES CENTRE-CUHAS BUGANDO Not for loan 20240610171113.0
Total holds: 0

Abstract :

The aim of this study was to determine the entrance surface air kerma (ESAK) in adult patients during digital radiography and to evaluate the optimisation potential in five common X-ray examinations in Tanzania. Based on a sample of 240–610 patients, ESAK was estimated using X-ray tube output measurements, patient information and backscatter factors. The results show that the mean ESAK values were higher or comparable to data from the literature. The diagnostic reference values of ESAK for digital radiography were 0.31 mGy (chest PA), 4 mGy (lumbar spine AP), 5.4 mGy (lumbar spine LAT), 3.8 mGy (abdomen AP) and 2.4 mGy (pelvis AP). For computed radiography, the mean ESAK ranges were 0.44–0.57 mGy (thoracic AP), 3.59–3.72 mGy (lumbar spine AP), 6.16–6.35 mGy (lumbar spine LAT), 3.89–3.44 mGy (abdominal AP) and 2.92–3.47 mGy (pelvic AP). In conclusion, high ESAK variations show the potential for optimising protection in digital radiology.